The CI sensor provides the advantages of infrared temperature measurement in a compact, low cost sensor and can easily replace traditional contact probes with a J-type or K-type thermocouple output, or with a 0-5V output, when integrated into a standard 4-wire system.
The CI sensor provides the advantages of infrared temperature measurement in a compact, low cost sensor and can easily replace traditional contact probes with a J-type or K-type thermocouple output, or with a 0-5V output, when integrated into a standard 4-wire system.
Discontinued!
This product has been discontinued and is no longer available. There is no direct replacement. Don't worry, we have some options to help you find what you need:
Visit our CI Series Compact IR Sensors category for similar products.
Or let's do a search and help you find what you need:
The unit's onboard electronics are protected by a rugged IP65 (NEMA-4) stainless steel housing, allowing the sensor to function in ambient temperatures to 160°F (70°C) without cooling. With water cooling, the sensor can withstand ambient temperatures to 500°F (260°C). Because the unit has the same 50 Ω output impedance as a thermocouple, it functions accurately - without offset errors - when used in conjunction with the thermocouple break protection circuitry in most controllers, displays, and transmitters.
Features
Applications
| Spectral Response | 7 to 18 μm |
| Temperature Resolution | <1°F or 0.5°C |
| Temperature Range | 32 to 932°F (0 to 500°C) |
| Optical Resolution | 4:1 |
| System Repeatability | ±1% of measured value or ±1°C (2°F), whichever is greater |
| Response Time (95%) | 350 mSec |
| Emissivity | Fixed at 0.95 |
| Outputs | User-selectable thermocouple output (model specific, either J or K) or voltage output 10 mV/°C |
| Cable Length | 3.2' (1 m) standard, longer cables optional |
| Output Impedance | 50 Ω |
| Minimum Load Impedance | 50 KΩ |
| Power Supply | 12 to 24 VDC (≤2.5% ripple) @ 20 mA |
| Environmental Rating | IP65 (NEMA-4) |
| Ambient Temperature Range | 32 to 160°F (0 to 70°C) With air cooling: 32 to 200°F (0 to 90°C) With water cooling: 32 to 500°F (0 to 260°C) |
| Storage Temperature | -22 to 185°F (-30 to 85°C) |
| Relative Humidity | 10 to 95%, non-condensing |
| Shock | IEC 68-2-27 (MIL STD 810D) 50 g's, 11 mSec, any axis |
| Vibration | IEC 68-2-27 (MIL STD 810D) 3 g's, any axis, 11–200 Hz |
| Weight | 4.5 oz (130 g) |
| Spectral Response | 7 to 18 μm |
| Temperature Resolution | <1°F or 0.5°C |
| Temperature Range | 32 to 932°F (0 to 500°C) |
| Optical Resolution | 4:1 |
| System Repeatability | ±1% of measured value or ±1°C (2°F), whichever is greater |
| Response Time (95%) | 350 mSec |
| Emissivity | Fixed at 0.95 |
| Outputs | User-selectable thermocouple output (model specific, either J or K) or voltage output 10 mV/°C |
| Cable Length | 3.2' (1 m) standard, longer cables optional |
| Output Impedance | 50 Ω |
| Minimum Load Impedance | 50 KΩ |
| Power Supply | 12 to 24 VDC (≤2.5% ripple) @ 20 mA |
| Environmental Rating | IP65 (NEMA-4) |
| Ambient Temperature Range | 32 to 160°F (0 to 70°C) With air cooling: 32 to 200°F (0 to 90°C) With water cooling: 32 to 500°F (0 to 260°C) |
| Storage Temperature | -22 to 185°F (-30 to 85°C) |
| Relative Humidity | 10 to 95%, non-condensing |
| Shock | IEC 68-2-27 (MIL STD 810D) 50 g's, 11 mSec, any axis |
| Vibration | IEC 68-2-27 (MIL STD 810D) 3 g's, any axis, 11–200 Hz |
| Weight | 4.5 oz (130 g) |